one) Fill in the use document in time following running the glove box, including “user, water oxygen price, what to do”.
2) When changing gasoline cylinders in the glove box, the stress is established to the default value ± 3mbar.
three) The daily working force is set according to the operating ease and comfort. When not in use for a extended time, the strain is established to 1-6mbar.
4) Frequently examine the put on of the gloves, verify the oil stage of the vacuum pump (among the upstream and downstream markings), and insert the pump oil correctly.
five) Clear the drinking water probe after every two to three months to make sure accurate detection. (Also want to refer to acrylic glove box of the atmosphere in various user boxes)
six) If you usually use organic and natural solvents in the glove box, it is advised that you regularly change the organic solvent adsorbent material, not afterwards than one yr at the latest.
seven) Substitute the filter factor often, no later on than two many years.
8) In accordance to the genuine use of your glove box, choose to replace the vacuum pump oil cycle, such as black and yellow pump oil, drinking water or much more organic and natural solvents in the pump oil. The oil replacement cycle need to not exceed two several years at the most current.
nine) If a huge amount of organic and natural solvent is inevitably uncovered in the course of the experiment, the cycle requirements to be stopped for the experiment. Following the experiment, initial clean and then start off the cycle.
ten) When the surroundings in the box is toxic or damaging gasoline, the exhaust port of the vacuum pump, the exhaust port of the regeneration gas and the exhaust port of the cleaning valve must be exhausted to the outside with an exhaust pipe (when exhaust is not allowed to be outdoor, the exhaust gasoline treatment method should be done first).